ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 4 - Functional Checks 4-1
PRELIMINARY
Chapter 4
Functional Checks
Section 4-1
Overview
4-1-1 Purpose of Chapter 4
This chapter describes the functional checks that must be performed either as a part of the installation
procedure, or as required during servicing and periodic maintenance of the Vivid S5 N or Vivid S6 N
ultrasound scanner.
NOTE: Refer to Chapter 7 - Diagnostics/Troubleshooting for detailed descriptions of some of the
functional checks described in this chapter.
4-1-2 Specific Equipment Required
Empty media (CD-R or memory card)
ECG Pads
•ECG Cable:
- AHA/AMERICAS, P/N: 2256477
OR
- IEC / EUROPE, ASIA, ROW, P/N: 2256478
At least one probe (ideally you should check all the probes used on the system by the site.)
See Section 9-9 "Probes" on page 9-16 for an overview of available probes.
Table 4-1 Contents in Chapter 4
Section Description Page Number
4-1
Overview
4-1
4-2
General Procedures
4-2
4-3
Functional Checks
4-18
4-4
Image Testing
4-38
4-5
System Turnover Checklist
4-45
4-6
Site Log
4-46

Table of Contents

Related product manuals