ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-53
PRELIMINARY
8-3-7 Swivel and Up-Down Handle Replacement Procedure
8-3-7-1 Tools
Flat and Phillips screwdrivers as needed.
8-3-7-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-3-7-3 Swivel and Up-Down Handle Replacement Procedure
1) Remove the Operator Panel as described in the “Keyboard Assembly Removal Procedure” on
page 8-39.
2) Remove the Probe Shelf as described in the “Probe Shelf Removal Procedure” on page 8-52.
3.) Unscrew the protective metal plate (shown in Figure 8-66) and rotate it upwards 90 degrees.
Figure 8-66 Protective Metal Plate Rotated 90 Degrees
Swivel Cable
Up-down Cable