ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 10 - Care and Maintenance 10-9
PRELIMINARY
10-4-4 Cleaning
10-4-4-1 General Cleaning
Frequent and diligent cleaning of the Vivid S5 N or Vivid S6 N ultrasound unit reduces the risk of
spreading infection from person to person, and also helps to maintain a clean working environment.
CAUTION
When performing Cleaning Procedures, to prevent the risk of system damage, always observe
the following precautions:
Use only cleaning materials and solutions recommended in the procedures described in Table 10-7 on page 10-10.
• Do not use any solutions or products not listed in the Vivid S5 N or Vivid S6 N User Manual.
• Do not spray any liquid directly onto the system covers, LCD Display or keyboard!
• Do not allow any liquid to drip or seep into the system.
• Prior to cleaning, turn OFF power to the system.
WARNINGWARNING
Table 10-7 on page 10-10 provides details for general cleaning.
Follow all safety instructions for all chemicals used for cleaning or other purposes as
instructed by manufacturer. Make sure you read the MSDS sections before usage!