ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
3-78 Section 3-4 - Preparing for Installation
PRELIMINARY
3-5-5-3 Disconnecting the Ultrasound Unit from the Electrical Outlet
1) Turn OFF the scanner.
2) Turn OFF the circuit breaker at the rear of the system.
3) Unplug the mains power cable from the AC wall outlet socket.
4) Unplug the mains power cable connector from the AC input socket at the bottom rear of the
scanner.
CAUTION
WHENEVER DISCONNECTING THE Vivid S5 N or Vivid S6 N UNIT FROM THE ELECTRICAL
OUTLET, ALWAYS OBSERVE THE SAFETY PRECAUTIONS. FIRST UNPLUG THE MAINS
POWER CABLE FROM THE WALL OUTLET SOCKET, THEN FROM THE UNIT ITSELF.
REMOVE BY PULLING ON THE CABLE CONNECTOR - DO NOT PULL ON THE CABLE.