EasyManua.ls Logo

GE Vivid S5 N - 4-3-1-9 Visual Test

GE Vivid S5 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
4-32 Section 4-3 - Functional Checks
PRELIMINARY
4-3-1-9 Visual Test
5 Trackball to the Start button and press Set
The test commences.
When the PENCIL RX Test sequence is complete, the
View Test Log window opens automatically, listing the
test performed and showing whether it passed or failed
- see the example in Figure 7-57 on page 7-88.
6
Use the scroll arrows at the bottom of the Reports
window to scroll through the channels and view the
corresponding results.
Table 4-13 Visual Test
Step Task Expected Result(s)
1
Connect the 3S-RS probe.
2
Increase the gain to see the homogenous noise.
The probe ID is 3S-RS. No strikes or artefacts are
present.
Table 4-12 Front End Tests (cont’d)