ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 4 - Functional Checks 4-45
PRELIMINARY
Section 4-5
System Turnover Checklist
Before returning the Vivid S5 N or Vivid S6 N ultrasound scanner to regular operational use, the System
Configuration settings should be checked. In 2D-Mode, press Config on the keyboard and check the
settings in accordance with those listed or referenced in Table 4-25 "Software Configuration Checklist"
on page 4-45
4-5-1 Software Configuration Checks
Table 4-25 Software Configuration Checklist
Step Task Check
1
Check the date and time settings. Refer to ‘Date and Time Adjustments” on page 3-81.
2
Check the language settings.Refer to ‘Language Selection” on page 3-82.
3
Ensure that all of the software options are configured correctly. Refer to ‘Software
Options Configuration” on page 3-95.
4
Ensure that the hospital details are configured correctly. Refer to Table 3-2 on page 80.
6
Ensure that the measurement units are configured as required for the specific site and
country.Refer to ‘Units of Measure” on page 3-84.
7
Ensure that the connectivity network settings have not been modified and are configured
as required for the specific site. Refer to ‘Set the Scanner’s Network Information” on
page 3-110.
8
Ensure that the Monitor - Internal (and External, if required) is properly adjusted. Refer to
6-1-2 "LCD Display Adjustments and Calibration" on page 6-1