4-3-1-7 ECG Test
5
Select Solid Test. Not more than 1 missing pixel in R/G/B tests
6
Select Red
Make sure there are no more than 1 missing pixel in the Full
Screen Red or display in another color.
7
Select Green
Make sure there are no more than 1 missing pixel in the Full
Screen Green or display in another color.
8
Select Blue
Make sure there are no more than 1 missing pixel in the Full
Screen Blue or display in another color.
9 When done, press Close
The system will prompt: Did the test pass?
Respond accordingly.
EXTERNAL MONITOR TEST:
1
Connect the external monitor to the system.
2
Boot-up the System in Maintenance Mode.
3
Press Alt +Config to start the diagnostic
package,
4
From the diagnostic test tree, select Back End.
This will activate PC Doctor diagnostics.
5
Run System Diagnostics\Monitor Test
6
Select Combination Test.
7
Follow the on-screen instructions. PASS
8
When done, press Close.
The system will prompt that the test passed.
Table 4-11 ECG Test
Step Task Expected Result(s)
1
Boot-up the System in Maintenance Mode and load the
application by double-clicking the Echoloader icon on
the desktop.
2
Connect the ECG cable to the input on the ECG port on
the right side of the scanner.
Connect the ECG leads to yourself (as the simulated
patient). Refer to the instructions in the User Manual.
3
Press Alt + Config to enter into Diagnostics Mode
4
From the Diagnostic Test tree, select Back End.
5
Run the ECG Test.
Table 4-10 Monitor Test (cont’d)
Step Task Expected Result(s)