GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-142 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-6 ECG Module Assembly Replacement Procedure (cont’d)
When disconnecting the cable, it is necessary to press the cable connector latch down (Figure 8-173):
8) Remove the ECG Module Assembly from the system - see Figure 8-174 on page 8-142.
8-6-6-5 ECG Module Assembly Installation Procedure
NOTE: When performing the following installation procedure steps, it may be helpful to review the movie clip
previously referenced in the 8-6-6-4 "ECG Module Assembly Removal Procedure" on page 8-140.
1) Return a new ECG Module assembly (and attached fastening bracket) to the cabinet cage
assembly and lay it upside down (on top of the battery), as shown in Figure 8-173 on page 8-142.
2) Re-connect the ECG cable (seen in Figure 8-176) to the ECG Module assembly (push it until it
clicks into position).
3) Re-connect the USB cable to the ECG Module assembly - seen in Figure 8-176.
Figure 8-173 Disconnecting the ECG Cable
Figure 8-174 ECG Module Assembly Removed