GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-141
PRELIMINARY
8-6-6 ECG Module Assembly Replacement Procedure (cont’d)
4) Loosen the three screws on the fastening bracket, as indicated in Figure 8-171.
(Access to one of the screws is via the battery bracket, as shown).
5) Carefully slide the ECG module assembly and fastening bracket sideways to the right (to move out
of the fixed position) and carefully lift it upwards and tilt it over, as shown in Figure 8-172
.
6) Disconnect the USB cable from the USB connector on the ECG module
(refer to Figure 8-170 on page 8-140).
7) Disconnect the ECG cable from the ECG connector on the ECG module (refer to Figure 8-170 on
page 8-140).
Figure 8-171 Loosening Screws on the Fastening Bracket
Figure 8-172 Removing the ECG Module Assembly and Bracket