GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-44 Section 8-3 - Control Console Components Replacement Procedures
PRELIMINARY
8-3-5 Keyboard Interface Assembly Replacement Procedure
8-3-5-1 Tools
Use the appropriate flat and Phillips type screwdrivers and the appropriate Allen keys.
8-3-5-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-3-5-3 Keyboard Interface Assembly Removal Procedure
NOTE: Figure 8-54 shows a general view of the Keyboard Interface Assembly work area.
1) Remove the Keyboard Assembly as described in the Keyboard Assembly Removal Procedure” on
page 8-39.
2) Remove the Probe shelf as described in the “Probe Shelf Removal Procedure” on page 8-52.
Figure 8-54 Keyboard Interface Assembly - General View of Work Area
USB Cable
Data Cable
LCD Power Cable
Gas Spring Lever
Swivel Lever
DVI Cable
Up/Down Arm