ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-116 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-2 Battery Replacement Procedure
NOTE: This procedure is applicable to Vivid S6 N systems only.
8-6-2-1 Tools
Use the appropriate Phillips-type screwdrivers as indicated in the Battery replacement procedure.
8-6-2-2 Time Required
1 minute
8-6-2-3 Preparations
Shut down the Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on page 4-7.
8-6-2-4 Battery Removal Procedure
1) Remove the Vivid S6 N left and right side covers, as described in the “Left Side Cover Removal
Procedure” on page 8-6 and “Right Side Cover Removal Procedure” on page 8-5.
2) Remove the Vivid S6 N front cover, as described in the “Front Cover Removal Procedure” on page 8-
17.
3) Open and remove the cabinet cage assembly, as described in “Opening and Removing the Cabinet
Cage Assembly” on page 8-109.
4) Using a small Phillips screwdriver, loosen (but do not remove) the screw on the side of the battery
compartment (near the bottom of the cage - see Figure 8-144).
Figure 8-144 Loosening the Screw on the Battery Compartment