7-110 Section 7-5 - Monitoring Tests
PRELIMINARY
7-5-1 Accessing the Monitoring Diagnostic Test Options (cont’d)
• If errors were identified but the results were marginal, perform calibration, as described in
“DC Offset Calibration” on page 7-27, and then repeat the test.
• If errors were identified and the results were not marginal, verify that all system cables and
boards are connected (not misplaced) and functioning properly, as described in Chapter 5 -
Components and Function (Theory).
If the problem persists, this indicates that either the specific board you tested is faulty or that
another board that supports the operation of this board is faulty. Contact your local
GE HEALTHCARE field engineer for assistance.
10.)Trackball to the Exit button and press Set to close the View Test Log dialog box.
11.)Press the <Esc> button on the keyboard to close the Diagnostic Test Window.
7-5-1-7 Tx Voltages Test
NOTE: For a description of this test, refer to the information in “Tx Voltages Test” on page 7-16.
1.) In the Diagnostic Test window, trackball to Tx Voltages in the list of Monitoring tests available for
selection from the Diagnostic Test Tree (see Figure 7-75 "Tx Voltages Test" on page 7-110) and
press Set. The Tx Voltages Test name is displayed under Execute; the relevant system
components to be tested are highlighted light blue in the Schematic View (AC/DC, Battery, PS), as
shown below:
2.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
In order to access an individual board test, it is necessary to insert the service dongle from system boot-
up and enter the Service Password.
Figure 7-75 Tx Voltages Test