ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-133
PRELIMINARY
29.)Click OK.
The following message is displayed:
30.)Click OK to close the message box.
NOTE: The DB links change according to the new system serial number to enable proper link
functionality.
31.)Reboot the system (without the service dongle) into normal working mode.
32.)From the Patient List, select an “old” Exam and make sure the images are displayed (and have
pointers to a Disk management device).
NOTE: If images are not displayed, do not proceed - contact OLC for assistance.
33.)Proceed to perform the following functionality tests:
- “Grounding Continuity” on page 10-23
- “Chassis Current Leakage Test” on page 10-24
- “Calibration Tests” on page 7-27
- “Full System Test” on page 7-11
Connectivity Settings
It is also be necessary to refer to the connectivity settings noted in step 5 of the Preparation procedure
(see “Preparation” on page 8-209) with regard to the following:
- TCP/IP address
- Network printer (if present)
- Wireless network settings (if present)
For step-by-step instructions, refer to “Connectivity Configuration” on page 3-109.
NOTE: All connectivity settings have to be configured with the co-operation of the local IT person.
Application Presets
Customer presets will remain as they were, since they are located in the original HDD. However the
factory presets may be different if newer-version software has been installed.
When restoring the Connectivity settings, it is necessary to perform the setup procedure through
Advanced Settings.