ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 5 - Components and Function (Theory) 5-13
PRELIMINARY
5-3-3 RFI_AM Board (cont’d)
5-3-3-7 Probe Management
5-3-3-7-1 Mechanism
The RFI Board performs the following probe management functions:
Controls the selection of a probe connector
Senses the Probe type
Senses the Scan Plane Angle on TE probes
Senses the Probe temperature
If the temperature exceeds the maximum temperature, the RBI stops scanning and disables
Tx voltage.
The temperature references are set to 41.3
o
C and 43.0
o
C
If the temperature reading falls below minimum temperature, the RBI stops scanning and disables
Tx voltage.
The minimum temperature references are set to 15
o
C. A temperature reading as low as 15
o
C
indicates a probe temperature-sensing error.
Table 5-3 RFI_AM Board - Input DC Voltages
Input Description Connector - PIN# Connected From
+3.3 V DC voltage from PS P14 1-4 From PS
HVP_IN Fixed DC high voltage +80 V P14-9 From PS
HVN_IN Fixed DC high voltage -80 V P14-11 From PS
+THV Transmit high voltage
varies from 0 to +50 V
P14-12 From PS
-THV Transmit high voltage
varies from 0 to -50 V
P14-14 From PS
+5 V_IN DC voltage from PS P14-16 From PS
-5 V_IN DC voltage from PS P14-20 From PS
+12 V_IN DC voltage from PS P14-18 From PS
1.8 V DC voltage from PS P1421, 22 From PS