ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-134 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-5 BEP Replacement Procedure
NOTE: It is recommended to replace the battery once every 2 years.
8-6-5-1 Tools
Use the appropriate Phillips and flat screwdrivers, as indicated in the BEP replacement procedure.
8-6-5-2 Time Required
15 minutes
8-6-5-3 Preparations
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
CAUTION
BEFORE REPLACING THE BEP, REMEMBER TO PERFORM A FULL INCREMENTAL
BACK-UP OF THE ARCHIVE AND USER PRESETS
.
For details, refer to the “Software Upgrade Procedure Overview” on page 8-207.
NOTICE
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to “Battery Removal Procedure” on page 8-116.