GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-251
PRELIMINARY
8-9-6-5 Peripherals Console Installation Procedure
1. Remove the DVD/CD-RW drive securing bracket from the new peripherals console
as shown in Figure 8-287.
2. Place the new peripherals console in its position on the Vivid S5/S6 system and tighten the securing
screws (previously released) - refer to Figure 8-286.
Note: The new peripherals console is supplied with two additional gaskets, which insulate the
B&W printer and hold the printer securely in the peripheral console. The gaskets are
adhered to either side of the printer as shown in Figure 8-288.
NOTE: The existing front gaskets are in place to close the gap between the front cover and the printer.
Figure 8-287 New Peripherals Console with DVD/CD-RW Drive Securing Bracket
Figure 8-288 B&W Printer showing the Additional Gasket Adhered
DVD/CD-RW Drive
Securing Bracket
Black and white
printer
Additional gasket
adhered to printer
adjacent rear
securing screw
(Second gasket on opposite side not visible in picture)
Air vents
Existing front
gaskets - installed
on either side of
printer