8-140 Section 8-6 - Lower Section Components Replacement Procedures
PRELIMINARY
8-6-6 ECG Module Assembly Replacement Procedure
8-6-6-1 Tools
Use the appropriate flat screwdriver, as indicated in the ECG Module Assembly replacement procedure.
8-6-6-2 Time Required
10 minutes
8-6-6-3 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-6-6-4 ECG Module Assembly Removal Procedure
1) Remove the system left and right side covers, as described in the 8-2-3-3 "Left Side Cover Removal
Procedure" on page 8-6 and 8-2-2-3 "Right Side Cover Removal Procedure" on page 8-5.
2) Remove the system front cover, as described in 8-2-6-3 "Front Cover Removal Procedure" on page
8-17.
3) Open and remove the cabinet cage assembly and place it on a stable surface, as described in 8-6-
1 "Opening and Removing the Cabinet Cage Assembly" on page 8-109.
NOTE: The location of the ECG Module Assembly is shown in Figure 8-170
.
NOTICE
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to “Battery Removal Procedure” on page 8-116.
Figure 8-170 ECG Module Assembly
ECG cable
USB cable
ECG module