GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-244 Section 8-8 - Software Loading
PRELIMINARY
NOTE: The total RAID 1 capacity is half the physical capacity of the two drives.
5) Connect the power cord from the power supply to the mains AC outlet.
6) Switch ON the power.
7) Study the status light indicators and verify that they function as described in Figure 8-278.
Figure 8-278 External USB Hard Disk - Status Light Indicators
8-9-4-6 Changing the Volume Name
NOTE: It is important that the External USB Hard Disk has a unique volume name. The site should choose a
naming convention to use for marking this Hard Drive and any future hard drives, in case more than one
drive will be needed in the future.
NOTE: Only connect one USB storage device (like this External USB Hard Disk) to the Vivid S5 N / Vivid S6 N
scanner at a time.
Follow these steps to set the Volume Name:
1) Connect the USB cable from the External USB Hard Disk to a laptop or PC.
2) Power up the External USB Hard Disk.
3) Under My Computer, after a while, the disk will show as a new drive, named Removable Disk
- refer to Figure 8-279 on page 8-245. (It will be allocated the next available drive letter.)
CHANGING THE VOLUME CONFIGURATION WILL ERASE ALL EXISTING DATA!
IF THE DRIVE CONTAINS FILES YOU WANT TO SAVE, MOVE THEM TO ANOTHER DRIVE
BEFORE CHANGING THE SWITCH SETTINGS.
1. HDD1 connection/power indicator
Glows white when the hard disk is connected and
the drive is powered up. Blinks when building a
mirrored RAID if HDD1 is the target disk.
2. HDD1 status indicator
Displays the status of the hard disk.
See the status indicator color/state chart below.
3. HDD2 connection/power indicator
Glows white when the hard disk is connected and
the drive is powered up. Blinks when building a
mirrored RAID if HDD2 is the target disk.
4. HDD2 status indicator
Displays the status of the hard disk. See the status
indicator color/state chart below.
Status Indicator Color and State:
• Flashing blue - Hard drive is being accessed
• Solid blue - Array is being rebuilt
• Red - Disk failure