P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-67
PRELIMINARY
11.)Using a screwdriver, tap out the supporting pins to release the Joint Assembly - Figure 8-83.
12.)Release the communication cables from the arm.
13.)Remove the cables by passing them through the Joint Assembly aperture.
14.)Remove the Joint Assembly.
8-4-3-4 Joint Assembly Installation Procedure
1.) Return the Joint Assembly to its location on the Upper Arm and insert the supporting pins.
2.) Secure the support pins by tightening the two locking screws - Figure 8-81.
3.) Return the ring clips to their positions on the supporting pins.
4.) Return the communication cables by passing them through the Joint Assembly aperture.
5.) Return the gas spring cable to its position (see Figure 8-80) with its bracket - the slot in the bracket
must face upwards - Figure 8-84.
6.) Reconnect the Ground lead.
7.) Return the Keyboard Interface Assembly to its position as described in the “Keyboard Interface
Assembly Installation Procedure” on page 8-47.
8.) Install the Lower Arm as described in the “Lower Arm Installation Procedure” on page 8-196.
9.) Install the Probe Shelf as described in the Probe Shelf Installation Procedure” on page 8-52
10.)Install the Keyboard Assembly as described in the “Keyboard Assembly Installation Procedure” on
page 8-42.
11.)Return the Monitor to its position as described in the “Monitor Installation Procedure” on page 8-29.
Figure 8-83 Removing the Joint Assembly
Figure 8-84 Gas Spring Cable Bracket showing Cable Slot
Tap out
Supporting
Pin
Slot in Bracket Facing Upwards