ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
8-100 Section 8-3 - Control Console Components Replacement Procedures
PRELIMINARY
8-5-1-3-3 Removing Cables from the Upper Arm Assembly
Note: After removing cables from the Keyboard Interface Assembly, it is recommended to lean
the keyboard interface platform backwards and to place protective material between the
lower arm and the handle, as indicated in Figure 8-117. It is also advisable to temporarily
tie the lower arm and handle together in this position, using string or a long tie wrap, as
appropriate.
1.) From below the Keyboard Interface Assembly, remove the two upper cable securing brackets, as
indicated in Figure 8-118.
(Use a Phillips screwdriver to carefully loosen the screw and remove it, together with the bracket
).
Figure 8-117 Positioning the Keyboard Interface Platform
Figure 8-118 Upper Cable Securing Brackets
Place Protective
Material Here
Bracket