P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-61
PRELIMINARY
8-4-2 Upper Arm Replacement Procedure
8-4-2-1 Tools
Use the appropriate flat and Phillips type screwdrivers and the appropriate Allen keys.
8-4-2-2 Preparation
Shut down the Vivid S5 N or Vivid S6 N ultrasound unit, as described in 4-2-3 "Power Shut Down" on
page 4-7.
8-4-2-3 Upper Arm Removal Procedure
1) If the rear tray is attached to the system, remove it before starting the procedure.
2) Remove the Monitor as described in the “Monitor Removal Procedure” on page 8-26.
3) Remove the Keyboard Interface Assembly as described in the “Keyboard Interface Assembly
Removal Procedure” on page 8-44.
4.) Remove the Lower Arm Assembly as described in the “Lower Arm Removal Procedure” on page 8-
193.
5.) Remove the Joint Assembly as described in the “Joint Assembly Removal Procedure” on page 8-
65.
6.) Remove the Gas Spring as described in the “Gas Spring Removal Procedure” on page 8-198.
7.) Remove the cables from the Upper Arm.
8.) Release the locking screw securing the Upper Arm support pin as shown in Figure 8-75.
Figure 8-75 Locking Screw Securing Support Pin on Lower Section of Upper Arm
Locking Screw
Support Pin
Upper Arm Assembly