ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
PRELIMINARY
Section 10-3
Maintenance Task Schedule
10-3-1 How Often Should Care & Maintenance Procedures be Performed?
The Customer Care & Maintenance Task Schedule (provided in Table 10-2 on page 10-5) specifies how
often the Vivid S5 N or Vivid S6 N ultrasound scanner should be serviced, and outlines items requiring
special attention.
NOTE: It is the customer’s responsibility to ensure the Care and Maintenance procedures are performed on the
Vivid S5 N or Vivid S6 N ultrasound scanner as scheduled in order to retain the high levels of safety,
dependability, and system performance.
Your GE HEALTHCARE Service Representative has an in-depth knowledge of your Vivid S5 N or Vivid
S6 N ultrasound scanning system and can best provide competent, efficient service. Please contact us
for details of your nearest GE HEALTHCARE Service Representative, and/or a quotation for services
by our qualified professional FSEs.
The service procedures and recommended intervals shown in the Customer Care & Maintenance Task
Schedule assumes that you use your Vivid S5 N or Vivid S6 N scanner for an average patient load
(10-12 patients per day) and that you do not use it as a primary mobile unit which is transported between
diagnostic facilities.
NOTE: If conditions exist which exceed typical usage and patient load, it is strongly recommended to increase
the maintenance frequencies.
Table 10-2 on page 10-5 shows the Customer Care & Maintenance Task Schedule.
Make sure the following are examined on a monthly basis:
• Connectors on cables - for any mechanical defects.
• The entire length of electrical and power cables - for cuts or abrasions.
• Equipment - for loose or missing hardware.
• The keyboard - for defects.
• Brakes - for defective operation.
TO AVOID ELECTRICAL SHOCK HAZARD, DO NOT REMOVE THE
PANELS OR COVERS FROM THE MACHINE. THIS SERVICING MUST
BE PERFORMED BY QUALIFIED SERVICE PERSONNEL ONLY.
FAILURE TO COMPLY, COULD RESULT IN SERIOUS INJURY!.
IF ANY DEFECTS ARE OBSERVED OR ANY SYSTEM MALFUNCTIONS OCCUR, DO
NOT OPERATE THE EQUIPMENT. INFORM A QUALIFIED SERVICE ENGINEER
IMMEDIATELY
.