EasyManua.ls Logo

GE Vivid S5 N - Page 643

GE Vivid S5 N
834 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-117
PRELIMINARY
8-6-2 Battery Replacement Procedure (cont’d)
5) Remove the metal securing bracket (with the screw attached) by sliding it to the side and lifting it up.
6) Slide the Battery upwards and out of the battery compartment along the grooves - until it is fully
removed from the system (see Figure 8-146).
NOTE:
WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF
IN ACCORDANCE WITH LOCAL REGULATIONS. ALTERNATIVELY, FORWARD IT
TO
GE HEALTHCARE FOR PROPER DISPOSAL.
Figure 8-145 Removing the Battery Securing Bracket
Figure 8-146 Sliding the Vivid S6 N Battery out of the Battery Compartment
Battery Securing Bracket

Table of Contents

Related product manuals