ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-223
PRELIMINARY
10. When prompted, type the following password:
geulsservice and press <Enter>.
Note: when typing the password, use lower-case letters.
A message is displayed showing the current system ID; the installation begins.
The CD then begins the software installation process and a progress indicator is
displayed, as shown below:
This stage of the process takes approximately 5 mins.
At the end of the process, the CD ejects and a message is then displayed
prompting you to press <CTRL + ALT + DELETE> to reboot the system.
11. Press <CTRL+ALT+DELETE>. The system re-boots.
12. Follow the instructions on the screen, clicking OK (or
Cancel) when prompted.
The machine performs a number of reboots before completing the installation.
As the installation proceeds, the system displays various installation screens.
The installation process will take approximately 15 minutes - DO NOT ATTEMPT
TO STOP THE INSTALLATION BEFORE COMPLETION!
13. If the PMC Firmware upgrade has not been updated, the following PMC
message is displayed:
14. Disconnect all peripherals from the USB and PCMCIA.
Move cursor over the displayed window and press the
SET button.
The following message is displayed
Table 8-3 Vivid™ S5 N and Vivid™ S6 N Software Re-Installation (cont’d)
Step Instruction Expected Result