EasyManua.ls Logo

HP 8560A

HP 8560A
450 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A9 Input Attenuator
Removal
1. Disconnect W34 at A15A2J1, and place the analyzer upside-down on the work bench.
2. Remove W41 and disconnect W44 from the attenuator.
3. Remove screw (1) securing the attenuator to the front-frame center support. See
Figure
3-
15.
4. Remove screw (1) securing the A9 Input Attenuator to the right side frame. See
Figure 3-16.
5. Remove the attenuator and disconnect the attenuator ribbon cable.
Replacement
1. Connect the attenuator-control ribbon cable to the A9 Input Attenuator.
2. Place the A9 Input Attenuator into the analyzer with the A9 mounting brackets resting
against the front-frame center support and the right side frame. Use caution to avoid
damaging any cables.
3. Attach the attenuator to the center support with one
panhead
screw (1). See Figure 3-15.
4. Attach the attenuator to the right side frame, using one
flathead
screw (1). See
Figure
3-
16.
5. Connect semirigid cables
W4l
and W44 to the attenuator assembly. Connect opposite end
of W41 to the front frame.
I
0
ii
0
Figure 3-18. A9 Mounting Screws at Right Frame
3-30 Assembly Replacement

Table of Contents

Related product manuals