Error Message Elimination
When an error message is displayed, always perform the following procedure.
1. Press
ISAVE]
and
SAVE STATE.
2. Store the current state in a convenient STATE register. (It may be necessary to set
SAVELOCK
to
OFF.)
3.
Press
ICAL)
and REALIGN LO
%
IF . Wait for the sequence to finish.
4.
Press
(RECALL)
and
RECALL STATE.
5. Recall the previously stored STATE.
6. If an error message is still displayed, refer to the list of error messages below for an
explanation of the error messages.
System Analyzer Programming Errors (100 to 150)
Refer to the HP
856OA/856lB/8563A
Spectrum Analyzer Operating and Programming Manual
for information on programming the analyzer.
100 NO PWRON
Power-on state is invalid; default state is loaded. Press (SAVE),
PWR ON STATE
to clear error message.
101
NO STATE
106
ABORTED!
107
HELLO ??
108
TIME OUT
109
CtrlFail
110.
NOT CTRL
111
# ARGMTS
112
??CMD??
113
FREQ NO!
114
TIME NO!
115
AMPL NO!
116
?UNITS??
117
NOP NUM
118
NOP EP
119
NOP UPDN
120
NOP
ONOF
121
NOP ARG
122
NOP TRC
123
NOP ABLK
State to be
RECALLed
not valid or not SAVEd.
Current operation is aborted; HP-IB parser reset.
No HP-IB listener is present.
Analyzer timed out when acting as controller.
Analyzer unable to take control of the bus.
Analyzer is not system controller.
Command does not have enough arguments.
Unrecognized command.
Command cannot have frequency units.
Command cannot have time units.
Command cannot have amplitude units.
Unrecognizable units.
Command cannot have numeric units.
Enable parameter cannot be used.
UP/DN
are not valid arguments for this command.
ON/OFF are not valid arguments for this command.
AUTO/MAN are not valid arguments for this command.
Trace registers are not valid for this command.
A-block format not valid here.
6-16 General Troubleshooting