124
NOP IBLK
125
NOP STRNG
126
NO ?
127
BAD DTMD
128
PK WHAT?
129
PRE TERM
130
BAD TDF
131
?? AM/FM
132
!FAV/RMP
133
!INT/EXT
134
??? ZERO
135
??? CURR
136
??? FULL
137
??? LAST
138
!GRT/DSP
'
139
PLOTONLY
‘
140 ?? PWRON”
141
BAD ARG
142
BAD ARG
143
NO PRESL
144
COUPL??
I-block format not valid here.
Strings are not valid for this command.
This command cannot be queried.
Not a valid peak detector mode.
Not a valid peak search parameter.
Premature A-block termination.
Arguments are only for TDF command.
AM/FM are not valid arguments for this command.
FAV/RAMP are not valid arguments for this command.
INT/EXT are not valid arguments for this command.
ZERO is not a valid argument for this command.
CURR is not a valid argument for this command.
FULL is not a valid argument for this command.
LAST is not a valid argument for this command.
GRT/DSP are not valid arguments for this command.
Argument can only be used with PLOT command.
PWRON is not a valid argument for this command.
Argument can only be used with FDIAG command.
Query expected for FDIAG command.
No preselector hardware to use command with.
Invalid COUPLING argument, expected AC or DC.
ADC Errors (200 to 299)
200 SYSTEM
201 SYSTEM
250
OUTOF
RG
251 NO
IRQ
These errors are directly related to the ADC/Interface Section.
Suspect a faulty A2 Controller or A3 Interface assembly.
ADC
Driver/ADC
Hardware/firmware interaction; check for other
errors.
ADC
Controller/ADC
Hardware/firmware interaction; check for other
errors.
ADC input is outside of ADC range.
Microprocessor not receiving interrupt from ADC.
General Troubleshooting 6-17