Table 1-4. Recommended Test Equipment
Instrument
Synthesized
Sweeper
(two required)
Critical Specifications
for Equipment Substitution
Frequency Range: 10 MHz to 7
GHz
Frequency Accuracy (CW): 1 x
10eg/day
Leveling Modes: Internal
&
External
Modulation Modes: AM
&
Pulse
Recommended Use
Model
HP
8340A*
PAT,
M,V
Synthesizer/
Level Generator
Power Level Range: -80 to
+16
dBm
Frequency Range: 200 Hz to 80 MHz
Frequency Accuracy: 1 x
10e7/month
HP
3335A*
PAT,
M,V
Flatness: f0.15 dB
Attenuator Accuracy: <&0.09 dB
Synthesized
Frequency Range: 5 MHz to 2.5
GHz
HP 8663A
p,v
Signal Generator Residual SSB Phase Noise at 10 kHz offset
(320 MHz <fc <640 MHz): <-131 dBc/Hz
Pulse/Function Frequency Range: 10 kHz to 50 MHz HP 8116A
P
Generator
Pulse Width: 200 ns;
Output Amplitude: 5 V peak-to-peak
Functions: Pulse
&
Triangle
TTL Sync Output
AM/FM
Frequency Range: 1 MHz to 200 MHz HP 8640B
A
Signal Generator
Frequency Modulation Mode
Modulation Oscillator Frequency: 1
kHz
FM Peak Deviation: 5 kHz
Microwave Frequency Range: 9 MHz to 6.9 GHz
HP
5343A*
P,A,M,V
Frequency Counter Time base Accuracy (Aging):
<5
x
lo-“/day
of
Option 001
External Frequency Reference Input
Universal
Modes: TI
AFB, Frequency Count
HP
5334A/B
P
Counter
Time Interval Measurement Range: 20 ms to 75 s
Frequency Measurement Range: 17
kHz to 11 MHz
Frequency Resolution: 1 MHz at 10 MHz
External Frequency Reference Input
Time base accuracy (Aging):
<3
x
10m7/month
Oscilloscope
Bandwidth (3
dB): dc to 100 MHz
HP
1980A/B* A,T
Minimum Vertical Deflection Factor:
<2
mV/div
Spectrum Analyzer
Frequency Range: 4
kHz to 7 GHz
HP
8566A/B
PAT
Typical Residual FM:
<l
Hz Pk-to-Pk in 100 ms
(Fundamental Mixing)
* Part of Microwave Workstation
P
=
Performance Tests; A
=
Adjustments; M
=
Test
&
Adjustment Module; T = Troubleshooting;
V
=
Operation Verification
General Information
l-
13