510 AMPL
.3M
511 AMPL
IM
512 RBW 100
513 RBW 300
514 RBW
IK
515
RBW
3K
516
RBW
IOK
517
RBW
100
518 RBW
300
519 RBW
IK
520 RBW 3K
521 RBW
IOK
522 RBW
IOK
523 RBW
IOK
524 RBW
IOK
525 RBW
IOK
526 RBW
<300
527 RBW
<300
528 RBW
<300
529 RBW
<300
530 RBW
(300
Unable to adjust amplitude of 300 kHz Resolution Bandwidth.
Insufficient gain during LC BW Cal.
Unable to adjust amplitude of 1 MHz Resolution Bandwidth.
Insufficient gain during LC BW Cal.
Unable to adjust 100 Hz Resolution Bandwidth. Insufficient gain
during crystal BW Cal.
Unable to adjust 300 Hz Resolution Bandwidth. Insufficient gain
during crystal BW Cal.
Unable to adjust 1 kHz Resolution Bandwidth. Insufficient gain
during crystal BW Cal.
Unable to adjust 3 kHz Resolution Bandwidth. Insufficient gain
during crystal BW Cal.
Unable to adjust 10 kHz Resolution Bandwidth. Insufficient gain
during crystal BW Cal.
Unable to adjust 100 Hz Resolution Bandwidth. Crystal Sweep
problem.
Unable to adjust 300 Hz Resolution Bandwidth. Crystal Sweep
problem.
Unable to adjust 1 kHz Resolution Bandwidth. Crystal Sweep
problem.
Unable to adjust 3 kHz Resolution Bandwidth. Crystal Sweep
problem.
Unable to adjust 10 kHz Resolution Bandwidth. Crystal Sweep
problem.
Unable to adjust symmetry of 10 kHz Resolution Bandwidth in first
XTAL pole.
Unable to adjust symmetry of 10 kHz Resolution Bandwidth in second
XTAL pole.
Unable to adjust symmetry of 10 kHz Resolution Bandwidth in third
XTAL pole.
Unable to adjust symmetry of 10 kHz Resolution Bandwidth in Fourth
XTAL pole.
ADC timeout during IF ADJUST of <300 Hz Resolution Bandwidth.
Step gain correction failed for <300 Hz Resolution Bandwidth. Check
narrow BW SGO attenuator.
Calibration of dc level at ADC failed for <300 Hz Resolution
Bandwidth.
Invalid demodulated data for <300 Hz Resolution Bandwidth flatness
and IF down-converter. Demod data for calibration may be bad.
Adjustment of VCXO down-converter failed. Narrow BW VCXO
calibration failed.
General Troubleshooting 6-27