14. External Mixer Bias Adjustment
3. On the HP 8560A press
[m)
EXTERNAL MIXER BIAS, then BIAS OFF.
4. Adjust A15R926 EXT BIAS ZERO for a DVM reading of 0.000 Vdc
512.5
mV.
2-56 Adjustment Procedures