Pathway Technical Reference Manual
Proprietary 180 of 190
Sensor Heater\TEC: The number of
defined calibration table entries is
less than the required 15.
The calibration cannot be completed because
there are not enough calibrated points.
Calibrate at least 15 points in order to
complete the manual calibration.
The number of defined calibration
table entries is more than the
required 15, but less than the
recommended 20.
1. Click YES if you want PATHWAY to
perform interpolation and fill out the
missing values.
2. Click NO to enable return to the
Hardware Settings and fill in missing
values.
The recommended number of points for
calibration is 20. Select YES to complete the
manual calibration based on the existing
points (less than 20) or select NO to return
and calibrate more points.
Emergency button was pressed /
Hardware safety was triggered
The system detected that the emergency
button has been pressed, or the hardware
safety mechanism has been triggered. The
system is now Safe Mode in order to prevent
any damage.
1. Release the emergency button and \ or
the Emergency button on the extended
cable
2. Run a system Self-Test.
2.1. If the Self-Test completes without
any further errors the system can
return to normal operation.
2.2. If the Self-Test cannot be complete,
restart the system and retry. If the
Self-Test continues to fail, contact
Medoc Service for assistance.
During Self-Test MEDOC Main Station
detected that the hardware safety
was triggered.
1. Verify that Emergency Button is
released.
2. Click OK to restart Self-Test.
The system detected that the emergency
button has been pressed, or the hardware
safety mechanism has been triggered. Release
the emergency button and \ or the Emergency
button on the extended cable and restart the
Self-Test.
The test cannot be started because
Medoc Main Station is not connected
to any device or Demo Mode.
Medoc Main Station needs to be connected to
the PATHWAY system (or to the simulator in
Demo Mode) in order to start the test. Check
that there is communication between
PATHWAY and Medoc Main Station. Refer to
section 10.1.1 for further information.
During automatic calibration software
safety protection is lowered;
therefore, it is not allowed to place
Thermode on human subjects!
1. Check that the Thermode is
properly placed in the jig and
connected to the Pathway.
2. Click YES to continue calibrating.
3. Click NO to discontinue calibration
Warning notification before automatic
calibration.
Follow the displayed instructions, and
continue.