Monitoring Noninvasive Blood Pressure (NIBP) (Optional) Interactions after Pairing a TM80 with a BP10
TM80 Telemetry Monitor Operator’s Manual 10 - 9
DIA Configures whether to trigger
the NIBP diastolic pressure
alarm.
On, Off
Alarm limit range:
For adult:
11 mmHg to 209 mmHg; the default alarm
high limit is 90 and the default alarm low limit
is 50.
For pediatric:
11 mmHg to 149 mmHg; the default alarm
high limit is 70 and the default alarm low limit
is 40.
Alarm priority: Med, High
MAP Configures whether to trigger
the NIBP mean pressure alarm.
On, Off
Alarm limit range:
For adult:
21 mmHg to 229mmHg; the default alarm
high limit is 110 and the default alarm low
limit is 60.
For pediatric:
Alarm limit range: 21 mmHg to 164 mmHg;
the default alarm high limit is 90 and the
default alarm low limit is 50.
Alarm priority: Med, High
SYS
Extreme
Configures whether to trigger
the extreme NIBP systolic
pressure alarm.
On, Off
Alarm limit range:
For adult:
40 mmHg to 270 mmHg; the default alarm
high limit is 175 and the default alarm low
limit is 75.
For pediatric:
40 mmHg to 200 mmHg; the default alarm
high limit is 130 and the default alarm low
limit is 60.
Alarm priority: High
Options Description Settings*
* The factory default settings are in bold.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals