EasyManuals Logo

Mindray BeneVision TM80 User Manual

Mindray BeneVision TM80
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #140 background imageLoading...
Page #140 background image
Interactions after Pairing a TM80 with a BP10 Monitoring Noninvasive Blood Pressure (NIBP) (Optional)
10 - 10 TM80 Telemetry Monitor Operator’s Manual
DIA
Extreme
Configures whether to trigger
the extreme NIBP diastolic
pressure alarm.
On, Off
Alarm limit range:
For adult:
10 mmHg to 210 mmHg;
the default alarm high limit is 105 and the
default alarm low limit is 35.
For pediatric:
10 mmHg to 150 mmHg; the default alarm
high limit is 80 and the default alarm low limit
is 30.
Alarm priority: High
MAP
Extreme
Configures whether to trigger
the extreme NIBP mean
pressure alarm.
On, Off
Alarm limit range:
For adult:
20 mmHg to 230 mmHg;
the default alarm high limit is 125 and the
default alarm low limit is 45.
For pediatric:
20 mmHg to 165 mmHg; the default alarm
high limit is 100 and the default alarm low
limit is 40.
Alarm priority: High
Options Description Settings*
* The factory default settings are in bold.
NOTE
• For systolic pressure, diastolic pressure, or mean pressure, its alarm high
limit should not be greater than the high limit of SYS Extreme, DIA
Extreme, or MAP Extreme.
• For systolic pressure, diastolic pressure, or mean pressure, its alarm high
limit should beat least 5mmHg greater than its alarm low limit.

Table of Contents

Other manuals for Mindray BeneVision TM80

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision TM80 and is the answer not in the manual?

Mindray BeneVision TM80 Specifications

General IconGeneral
BrandMindray
ModelBeneVision TM80
CategoryMonitor
LanguageEnglish

Related product manuals