Interactions after Pairing a TM80 with a BP10 Monitoring Noninvasive Blood Pressure (NIBP) (Optional)
10 - 10 TM80 Telemetry Monitor Operator’s Manual
DIA
Extreme
Configures whether to trigger
the extreme NIBP diastolic
pressure alarm.
On, Off
Alarm limit range:
For adult:
10 mmHg to 210 mmHg;
the default alarm high limit is 105 and the
default alarm low limit is 35.
For pediatric:
10 mmHg to 150 mmHg; the default alarm
high limit is 80 and the default alarm low limit
is 30.
Alarm priority: High
MAP
Extreme
Configures whether to trigger
the extreme NIBP mean
pressure alarm.
On, Off
Alarm limit range:
For adult:
20 mmHg to 230 mmHg;
the default alarm high limit is 125 and the
default alarm low limit is 45.
For pediatric:
20 mmHg to 165 mmHg; the default alarm
high limit is 100 and the default alarm low
limit is 40.
Alarm priority: High
Options Description Settings*
* The factory default settings are in bold.
NOTE
For systolic pressure, diastolic pressure, or mean pressure, its alarm high
limit should not be greater than the high limit of SYS Extreme, DIA
Extreme, or MAP Extreme.
For systolic pressure, diastolic pressure, or mean pressure, its alarm high
limit should beat least 5mmHg greater than its alarm low limit.