Troubleshooting Physiological Alarm Messages at the TM80
TM80 Telemetry Monitor Operator’s Manual 15 - 5
ECG Asystole High The patient is in arrhyth-
mia.
Check the patient’s con-
dition and the ECG con-
nections.
V-Fib/V-Tach
V-Tach
Vent Brady
Extreme Tachy
Extreme Brady
R on T Med*
Run PVCs Low*
Couplet Prompt*
Multiform PVC Med*
PVC Prompt*
Bigeminy Med*
Trigeminy
Tachy
Brady
Missed Beats Prompt*
Nonsus V-Tach Med*
Vent Rhythm
Pause Low*
Irr Rhythm Prompt*
A-Fib
PVCs/min Med*
Pauses/min
Pacer Not Cap-
ture
Prompt* The pacer appears
abnormal.
Check the pacer.
Pacer Not Pac-
ing
Measurement Alarm message
Default
Alarm
priority
Possible cause Solution
* The asterisk (*) means the alarm priority is configurable.