Physiological Alarm Messages at the TM80 Troubleshooting
15 - 6 TM80 Telemetry Monitor Operator’s Manual
ECG Irr Rhythm End Prompt Irregular rhythm no lon-
ger detected for the 
irregular rhythm end 
delay time.
Check the patient’s con-
dition and the ECG con-
nections.
A-Fib End Prompt Atrial fibrillation no lon-
ger detected for the Afib 
end delay time.
ST Single 
Alarm 
Violation
Med* The ST value of any ST 
lead has risen above 
the high alarm limit or 
fallen below the low 
alarm limit.
Check the patient’s 
condition and check if 
the patient category 
and alarm limit 
settings are correct.
ST Dual Alarm 
Violation
Med* The ST value of two or 
more ST leads has 
risen above the high 
alarm limit or fallen 
below the low alarm 
limit.
Check the patient’s 
condition and check if 
the patient category 
and alarm limit 
settings are correct.
ST-XX** High Med* The ST value has risen 
above the high alarm 
limit or fallen below 
the low alarm limit.
Check the patient’s 
condition and check if 
the patient category 
and alarm limit 
settings are correct.
ST-XX** Low Med*
QTc  High Med* The QT value has risen 
above the high alarm 
limit.
Check the patient’s 
condition and check if 
the patient category 
and alarm limit 
settings are correct.
ΔQTc  High Med*
Measurement Alarm message
Default 
Alarm 
priority
Possible cause Solution
* The asterisk (*) means the alarm priority is configurable.
** XX is the ECG lead name.