Troubleshooting Technical Alarm Messages at the TM80
TM80 Telemetry Monitor Operator’s Manual 15 - 11
SpO
2
SpO
2
Low Per-
fusion
Prompt
The SpO
2
sensor is not
properly placed or the
patient’s perfusion index
is too low.
1. Check the sensor
and sensor position.
2. Reposition the
sensor if necessary.
PR Over-
range
Low A The measured PR value
exceeds the measure-
ment range.
Contact Mindray or
your service person-
nel.
NIBP NIBP Low
Battery
Med A The battery charge is low. Replace with a new
battery.
NIBP Bat-
tery
Depleted
High A The battery charge is
almost depleted.
NIBP
Voltage
Error
High A The lithium-ion battery
voltage is out of the spec-
ified range.
NIBP Bat-
tery Error
High A The lithium-ion battery
communication is error.
NIBP Bat-
tery
Mainte-
nance
Required
Med A The lithium-ion battery is
aging.
MPAN
Discon-
nected
Med
&
The MAPN is discon-
nected.
Enable the MPAN
switch.
NIBP
Clock
Needs To
Be Set
Low C The button cell does not
have sufficient charge.
Reset the system
time for the BP10.
Measurement
Alarm
message
Default
Alarm
priority
Alarm
Indication
Possible cause Solution
* The asterisk (*) means the alarm priority is configurable.
** XX is the ECG lead name.