Technical Alarm Messages at the TM80 Troubleshooting
15 - 12 TM80 Telemetry Monitor Operators Manual
NIBP NIBP
Module
Error
High A
ƾ An error occurred
to the NIBP
module.
ƾ There is a problem
with
communication
between the TM80
and BP10.
Restart the BP10.
NIBP Cuff
Loose
Low C
ƾ The NIBP cuff is
not properly
connected.
ƾ There is a leak in
the airway.
1. Check the
patient's con-
dition and
verify patient
category.
2. Replace with
an appropri-
ate cuff and
connect it
correctly.
NIBP
Airway
Error
Low C An airway error occurred. Check the airway.
NIBP
Weak
Signal
Low C The patient’s pulse is
weak or the cuff is loose.
Check the patient’s
condition and
change the cuff
application site. If
the error persists,
replace the cuff.
NIBP
Over-
range
Low C The measured NIBP value
is not within the specified
range.
Contact your service
personnel.
NIBP
Exces-
sive
Motion
Low C Patient’s arm moves too
much.
Check the patient’s
condition and
reduce the patient
motion.
NIBP Cuff
Over-
pressure
Low C The NIBP airway may be
occluded.
Check the airway
and measure again.
Measurement
Alarm
message
Default
Alarm
priority
Alarm
Indication
Possible cause Solution
* The asterisk (*) means the alarm priority is configurable.
** XX is the ECG lead name.