Troubleshooting Technical Alarm Messages at the TM80
TM80 Telemetry Monitor Operator’s Manual 15 - 13
NIBP NIBP Cuff
or Air-
way Leak
Low C The NIBP airway may leak
air.
1. Verify that the
cuff is prop-
erly con-
nected.
2. 2. Verify that
the airway
does not leak
air.
NIBP
Timeout
Low C
ƾ Time is out.
ƾ The measurement
time is over 120
seconds.
1. Check the
patient’s con-
dition and
NIBP connec-
tions.
2. Replace the
cuff.
NIBP Cuff
and
Patient
Mis-
match
Low C The cuff type applied
mismatches the patient
category.
1. Check the
patient’s
category.
2. Replace the
cuff.
Intervals
Not Set
Low C The interval in Sequence
mode is not set.
Set the intervals.
NIBP-S
Over-
range
Low
$
The measured NIBP value
is not within the mea-
surement range.
Contact your service
personnel.
NIBP-D
Over-
range
Low
$
NIBP-M
Over-
range
Low
$
Measurement
Alarm
message
Default
Alarm
priority
Alarm
Indication
Possible cause Solution
* The asterisk (*) means the alarm priority is configurable.
** XX is the ECG lead name.