EMC
TM80 Telemetry Monitor Operator’s Manual B - 3
RF emissions
CISPR 11
Class A The system is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies build-
ings used for domestic purposes.
Harmonic distortion
IEC 61000-3-2
Class A
Voltage fluctuations
and flicker IEC 61000-3-3
Compliance
Note: The TM80 Telemetry Monitor and BP10 (subpart) are classified by is their using environment or intended
function.
NOTE
The device needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information pro-
vided below.
Other devices may affect this device even though they meet the require-
ments of CISPR.
When the inputted signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.
The EMISSIONS characteristics of this device make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residen-
tial environment (for which CISPR 11 class B is normally required) this
device might not offer adequate protection to radio-frequency commu-
nication services. The user might need to take mitigation measures, such
as relocating or re-orienting the device.
If the essential performance is lost or degraded, it may be necessary to
take mitigation measures, such as re-orienting or relocating the ME
EQUIPMENT or ME SYSTEM or shielding the location or stopping using
the monitor and contact the service personnel.