EMC
TM80 Telemetry Monitor Operator’s Manual B - 3
RF emissions
CISPR 11
Class A The system is suitable for use in all establishments other 
than domestic and those directly connected to the public 
low-voltage power supply network that supplies build-
ings used for domestic purposes.
Harmonic distortion
IEC 61000-3-2
Class A
Voltage fluctuations
and flicker IEC 61000-3-3
Compliance
Note: The TM80 Telemetry Monitor and BP10 (subpart) are classified by is their using environment or intended 
function.
NOTE
• The device needs special precautions regarding EMC and needs to be 
installed and put into service according to the EMC information pro-
vided below.
• Other devices may affect this device even though they meet the require-
ments of CISPR.
• When the inputted signal is below the minimum amplitude provided in 
technical specifications, erroneous measurements could result.
• The EMISSIONS characteristics of this device make it suitable for use in 
industrial areas and hospitals (CISPR 11 class A). If it is used in a residen-
tial environment (for which CISPR 11 class B is normally required) this 
device might not offer adequate protection to radio-frequency commu-
nication services. The user might need to take mitigation measures, such 
as relocating or re-orienting the device.
• If the essential performance is lost or degraded, it may be necessary to 
take mitigation measures, such as re-orienting or relocating the ME 
EQUIPMENT or ME SYSTEM or shielding the location or stopping using 
the monitor and contact the service personnel.