EMC
B - 4 TM80 Telemetry Monitor Operator’s Manual
If the system is operated within the electromagnetic environment listed in Table Guid-
ance and declaration — electromagnetic immunity, the system will remain safe and pro-
vide the following essential performance: 
■ Operating mode 
■ Accuracy 
■ Function 
■ Accessories identification 
■ Data stored 
■ Alarm 
■ Detect for connection
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user 
of the device should assure that it is used in such an environment. 
Immunity test
IEC 60601 Test 
level
Compliance level
Electromagnetic environment - 
guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15kV air
±8 kV contact
±15kV air
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative humidity 
should be at least 30%.
Electrical fast 
transient/burst 
IEC 61000-4-4
±2 kV for power 
supply lines
±1 kV for input/out-
put lines
(length greater 
than 3 m)
±2 kV for power sup-
ply lines
±1 kV for input/out-
put lines
(length greater than 3 
m)
Mains power quality should be that of a 
typical commercial or hospital environ-
ment.
Surge
 IEC 61000-4-5
±1 kV line(s) to 
line(s)
±2 kV line(s) to 
earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Voltage dips 
and Voltage 
interruptions
IEC 61000-4-11
0 % U
T
 for 0,5 cycle
0 % U
T
 for 1 cycle 
and 70 % U
T
 for 25/
30 cycles
0 % U
T
 for 250/300 
cycle
0 % U
T
 for 0,5 cycle
0 % U
T
 for 1 cycle and 
70 % U
T
 for 25/30 
cycles
0 % U
T
 for 250/300 
cycle
Mains power quality should be that of a 
typical commercial or hospital environ-
ment. If the user of our product requires 
continued operation during power 
mains interruptions, it is recommended 
that our product be powered from an 
uninterruptible power supply or a bat-
tery.