Monitoring ECG, Arrhythmia, ST and QT Arrhythmia Monitoring
TM80 Telemetry Monitor Operator’s Manual 7 - 23
ǂ All OffὉswitches off all arrhythmia alarms. This button becomes inactive
when Lethal Arrhy Alarms Off is disabled in the System Setup menu of the
CMS.
Ç‚ Lethals Only: switches on all the lethal arrhythmia alarms only. Other types
of arrhythmia alarms are off.
5. In the Alarms section, select the options described in the following table.
Options Description Settings*
Asystole Configures arrhyth-
mia alarms.
On, Off
Asystole Delay: 3 seconds to 10 seconds
The default is 5 sec.
Alarm priority: High
V-Fib/V-Tach On, Off
Alarm priority: High
V-Tach On, Off
V-Tach PVCs: 3 beats to 99 beats; the default
is 6 beats.
V-Tach Rate:100 bpm~200 bpm; the default
is 130 bpm.
Alarm priority: High
* The factory default settings are in bold.