Arrhythmia Monitoring Monitoring ECG, Arrhythmia, ST and QT
7 - 24 TM80 Telemetry Monitor Operator’s Manual
Vent Brady Configures arrhyth-
mia alarms.
On, Off
V-Brady PVCs:3 beats to 99 beats; the default
is 5 beats.
V-Brady Rate: 15 bpm to 60 bpm; the default
is 40 bpm.
Alarm priority: High
Extreme Tachy On, Off
Extreme Tachy: 61 bpm to 300 bpm; the
default is 160 for adult and 180 for pediatric.
Alarm priority: High
Extreme Brady On, Off
Extreme Brady: 15 bpm to 119 bpm; the
default is 35 for adult and 50 for pediatric.
Alarm priority: High
R on T On, Off
Alarm priority: Prompt, Low, Med, High
Run PVCs On, Off
Alarm priority: Prompt, Low, Med, High
Couplet On, Off
Alarm priority: Prompt, Low, Med, High
Multiform PVC On, Off
Multif PVCs Window: 3 beats to 31 beats; the
default is 15 beats.
Alarm priority: Prompt, Low, Med, High
PVC On, Off
Alarm priority: Prompt, Low, Med, High
Options Description Settings*
* The factory default settings are in bold.