Monitoring ECG, Arrhythmia, ST and QT Arrhythmia Monitoring
TM80 Telemetry Monitor Operator’s Manual 7 - 25
Bigeminy Configures arrhyth-
mia alarms.
On, Off
Alarm priority: Prompt, Low, Med, High
Trigeminy On, Off
Alarm priority: Prompt, Low, Med, High
Tachy On, Off
60 bpm to 299 bpm; the default Tachy
threshold is consistent with the HR alarm
high limit.
Alarm priority: Med, High
Brady On, Off
16 bpm to 120 bpm; the default Brady
threshold is consistent with the HR alarm low
limit.
Alarm priority: Med, High
Pacer Not Cap-
ture
On, Off
Alarm priority: Prompt, Low, Med, High
Pacer Not Pacing On, Off
Alarm priority: Prompt, Low, Med, High
Missed Beats On, Off
Alarm priority: Prompt, Low, Med, High
Nonsus V-Tach On, Off
Alarm priority: Prompt, Low, Med, High
Vent Rhythm On, Off
Alarm priority: Prompt, Low, Med, High
Pause On, Off
Pause Time: 1.5s, 2.0s, 2.5s, 3.0s
Alarm priority: Prompt, Low, Med, High
Options Description Settings*
* The factory default settings are in bold.