Command reference
R&S
®
ZNL/ZNLE
1011User Manual 1178.5966.02 ─ 20
Parameters:
<Boolean> ON | OFF – enable or disable auto-configuration.
Example: See SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:
DEFine
Manual operation: See "Auto Config" on page 519
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
Queries the number of configured external power meters. The result is an integer num-
ber of power meters.
Example:
See SYSTem:COMMunicate:RDEVice:PMETer:CATalog?
Usage: Query only
Manual operation: See "Configured Devices" on page 517
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:DEFine <PmtrName>, <Driver>,
<Interface>, <Address>[, <SensorId>]
Configures an external power meter and adds it to the list of available power meters.
Suffix:
<Pmtr>
.
Number of the configured power meter. Power meters must be
numbered in ascending order, starting with 1. If a number is
reused, the previous power meter configuration is overwritten.
Power meters can be assigned several times so that the number
of configured power meters is practically unlimited.
Parameters:
<PmtrName> Name of the external power meter (string parameter). An empty
string means that no particular name is assigned to the power
meter.
<Driver> Power meter type (string parameter). The power meter type is
identical with the name of the power meter driver file (*.pwm)
stored in the resources\extdev subdirectory of the analyz-
er's program directory. Type the driver file name as shown in the
"Add External Powermeter" dialog, i.e. without the file extension
".pwm" (example: use 'NRVD', if the corresponding driver file
name is NRVD.pwm).
Alternative: The '<Driver>' string can also contain the power
meter driver file name with its complete path, e.g. 'C:\Program
Files\Rohde-Schwarz\Vector Network Analyzer\ZNL
Server\resources\extdev\nrvd.pwm'.
<Interface>
Interface type (string parameter): 'VXI-11', 'socket',
'GPIB0', ..., 'GPIB9', 'other'
<Address> Interface address (string parameter), depending on the interface
type. See Table 10-3
VNA command reference