VNA GUI reference
R&S
®
ZNL/ZNLE
516User Manual 1178.5966.02 ─ 20
Import File...
Imports the transmission coefficients from a trace file. The imported file must be either
in Touchstone (*.s<n>p) or in *.csv format; see also Chapter 9.4.2, "Trace files",
on page 237.
Remote command:
MMEMory:LOAD:CORRection:TCOefficient<Ch>
Recall... / Save...
You can save the displayed power loss list to a power meter correction list file with
extension (*.pmcl) and reload it in later sessions.
Remote command:
MMEMory:LOAD:CORRection:TCOefficient<Ch>
MMEMory:STORe:CORRection:TCOefficient<Ch>
10.12.3.4 External Power Meters dialog
The "External Power Meters" dialog configures external power meters so that they can
be used for measurements and power calibrations.
Background information
Refer to section Chapter 9.7.14, "Power sensor measurements (R&S ZNL only)",
on page 306.
The configuration of a new external power meter involves the following steps:
1. Connect the power meter to your VNA using a LAN (VXI-11), GPIB, or USB inter-
face.
2. If the power meter is connected via LAN, enable LAN Detection
Cal softtool