Command reference
R&S
®
ZNL/ZNLE
852User Manual 1178.5966.02 ─ 20
Example: See MMEMory:LOAD:CORRection
Usage: Setting only
Manual operation: See "Resolve Pool Link / Remove Pool Link" on page 524
MMEMory:LOAD:CORRection:TCOefficient<Ch> <TraceFile>[, <Trace>[,
<DeEmbedding Position>]]
Loads two-port transmission coefficients from the specified power meter correction file
or trace file to channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<DeEmbedding
Position>
BOTH | DUT | PM
Determines the two-port the command refers to:
DUT
The two-port between VNA and DUT (during measurement)
PM
The two-port between VNA and PM (during power calibration)
BOTH
Both positions
(parameter omitted)
If the parameter is omitted, the command refers to the two-port
selected using SOURce<Ch>:POWer<PhyPt>:CORRection:
TCOefficient:CALibration.
Setting parameters:
<TraceFile> String parameter specifying the name and directory of the loa-
ded trace file. The R&S ZNL/ZNLE supports power meter cor-
rection list files (*.pmcl, generated using MMEMory:STORe:
CORRection:TCOefficient<Ch>), *csv, and Touchstone
(*.s1p, *.s2p, ...) files.
If no path is specified the analyzer searches the current direc-
tory, to be queried with MMEMory:CDIRectory?
The file extensions *.s<n>p, *csv, and *.pmcl for Touch-
stone, ASCII, and power meter correction list files are manda-
tory.
*RST: n/a
VNA command reference

Table of Contents

Other manuals for R&S ZNL Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNL Series and is the answer not in the manual?

R&S ZNL Series Specifications

General IconGeneral
BrandR&S
ModelZNL Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals