VNA GUI reference
R&S
®
ZNL/ZNLE
518User Manual 1178.5966.02 ─ 20
The device driver does not match exactly the device type. Probably the device
cannot be controlled properly. Measurements results are questionable.
Remote command:
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:DEFine
SYSTem:COMMunicate:RDEVice:PMETer:DELete
SYSTem:COMMunicate:RDEVice:PMETer:CATalog?
SYSTem:COMMunicate:RDEVice:PMETer:COUNt?
Scan Instruments
Performs an autodetection of "Known Devices" on the supported interfaces.
Note: Unintentional switchover to remote control
When using the NI-VISA library, ensure that the network analyzer itself is not listed as
a network device in the Measurement & Automation Explorer. Otherwise, "Scan Instru-
ments" initiates an identification query (*IDN?), causing the analyzer to close the
"External Power Meters" dialog (without executing "Scan Instruments") and to activate
the remote screen.
Remote command:
n/a
LAN Detection
Activate "LAN Detection" to include the LAN interface in the autodetection sequence
(see Scan Instruments).
Note: "LAN detection" only works for external devices on the same IP subnet than your
VNA.
As a prerequisite, your VNA must accept incoming connections on UDP port 2473. If
necessary, add a corresponding inbound rule to your firewall settings.
Remote command:
n/a
Add Device
Adds a new instrument to the list of "Configured Devices".