VNA GUI reference
R&S
®
ZNL/ZNLE
527User Manual 1178.5966.02 ─ 20
CONFigure:TRACe<Trc>:CHANnel:NAME?
CONFigure:TRACe<Trc>:CHANnel:NAME:ID?
New Channel + Diagram
Creates a channel and a trace with default settings, and assigns the created trace to a
new diagram area.
The new channel is named Ch<n>, where <n> is the largest of all existing channel
numbers plus one. The name can be changed in the "Channel Manager" dialog.
Tip: To create a channel and a trace with default settings and to display the trace in the
active diagram area, use New Channel.
Copy Channel + Diagram
Copies the active channel with all its settings (including a possible channel calibration)
and traces, and displays the created traces in a new diagram area.
The new channel is named Ch<n>, where <n> is the largest of all existing channel
numbers plus one. The name can be changed in the "Channel Manager" dialog.
Tips: Use Copy Channel to display the new traces in an existing diagram area.
Delete Channel
Deletes the current channel including all the traces assigned to it and removes all dis-
play elements related to the channel from the diagram area. "Delete Channel" is dis-
abled if the channel setup contains only one channel: In manual control, each channel
setup must contain at least one diagram area with one channel and one trace.
Tips: Use the "Undo" function to restore a channel that was unintentionally deleted.
Remote command:
CONFigure:CHANnel<Ch>[:STATe] OFF
Channel On
Toggles the measurement state of the active channel.
Remote command:
CONFigure:CHANnel<Ch>:MEASure[:STATe]
Only Active Channel On
If checked, only the active channel is measured; all other channels are switched off
temporarily and dynamically.
Temporarily means: If "Only Active Channel On" is unchecked, then all channels return
to their original Channel On state.
Dynamically means: If another channel becomes the active one, the firmware adjusts
the temporary ON/OFF state accordingly.
Remote command:
CONFigure:CHANnel:MEASure:ACTive[:STATe]
Fixture Simulator
The "Fixture Simulator" switch deactivates or activates the configured deembedding,
embedding, balanced ports, and port impedance settings for the selected channel.
Channel softtool

Table of Contents

Other manuals for R&S ZNL Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ZNL Series and is the answer not in the manual?

R&S ZNL Series Specifications

General IconGeneral
BrandR&S
ModelZNL Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals