Command reference
R&S
®
ZNL/ZNLE
803User Manual 1178.5966.02 ─ 20
Manual operation: See "Channel table" on page 528
CONFigure:CHANnel:MEASure:ACTive[:STATe] <Boolean>
If set to ON (1), only the active channel is measured. All other channels are switched
off temporarily, i.e. until CONFigure:CHANnel:MEASure:ACTive is set to OFF.
Parameters:
<Boolean>
ON (1) | OFF (0)
Manual operation: See "Only Active Channel On" on page 527
CONFigure:CHANnel:MEASure:ALL[:STATe]
<Boolean>
Enables or disables the sweep in all channels of the active channel setup. This com-
mand can be used in combination with CONFigure:CHANnel<Ch>:MEASure[:
STATe] to optimize the measurement speed.
Parameters:
<Boolean> ON | OFF
*RST: ON
Example: See CONFigure:CHANnel<Ch>:MEASure[:STATe]
Manual operation: See "Continuous / Single" on page 343
CONFigure:CHANnel:MEASure:OPTimized <Order>[, <Ch1>, <Ch2>, ...]
This setting tells the firmware to optimize the channel switching times for a particular
channel sweep order.
Note that the optimization can only be applied, if all channels in the current setup are
in single sweep mode (INITiate<Ch>:CONTinuous OFF, and if there is no ongoing
sweep sequence. Any change in the current setup or a wrong sweep sequence lets the
analyzer fall back to unoptimized channel switching.
Setting parameters:
<Order> AUTO | MANual
AUTO
Optimize the switching times for the channel order that is auto-
matically used with INIT:ALL (increasing order of channel
numbers).
MANual
Optimize the switching times for the channel order manually
specified using channel numbers <Ch1>, <Ch2>, ...
<Ch1>, <Ch2>, ... Can only and must be specified for MANual optimized channel
ordering. Defines the order in which the channels are measured.
To get the optimization to work, you have to sweep the channels
manually, and in the specified order: INIT<Ch1>,
INIT<Ch2>, ...