ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 1001
Test port number of the analyzer. This suffix is ignored because the selected
power sweep range applies to all source ports used in the active channel.
–39.99 dBm to +10 dBm. The exact range depends on the analyzer model;
refer to the data sheet [dBm]. UP and DOWN increment/decrement the source
power in 1-dB steps.
Confirmed, command or query
SWE:TYPE POW
Activate a power sweep.
SOUR:POW:STAR -6; STOP 10
Select a power sweep range between –6 dBm and +10 dBm.
If the stop power entered is smaller than the current start power (SOURce<Ch>:POWer<Pt>:STARt),
the start power is set to the stop power minus the minimum power span of 0.01 dB.
ALC
SOURce<Ch>:POWer<Pt>:ALC...
This subsystem controls the port and channel specific Automatic Level Control (ALC) settings.
SOURce<Ch>:POWer<Pt>:ALC:AUBW <Boolean>
Enables or disables automatic bandwidth setting for port <Pt> in channel <Ch> .
ON - automatic bandwidth selection (ALC Path IF Bandwidth in ALC Config
dialog set to Auto)
OFF - manual bandwidth selection (using
SOURce<Ch>:POWer<Pt>:ALC:BANDwidth)
Device-specific, command or query.
SOURce1:POWer:ALC:CSTAate ON
Allow automatic level control for channel 1.
SOURce1:POWer1:ALC ON; SOURce1:POWer2:ALC ON
Enable automatic level control for ports 1 and 2 in channel 1.